Wash-Checks
Chemical control system to monitor the cleaning achieved with washer-disinfectors during daily routine usage.
Intended purpose
Easy too use control system to monitor the cleaning efficiency of medical instruments in washer-disinfectors.
| Product | Order Nr. | Package size | Storage | Durability |
| Wash-Checks retainer | 830001 | 1 Piece | dark and dryat room temperature | 2 years startingfrom date of manufacture |
| Wash-Checks | 830010 | 10 Pieces | ||
| indicator leaflets | 830050 | 50 Pieces | ||
| 830100 | 100 Pieces |
Application/ Process
Insert the indicator leaflet laterally into the retainer, making certain that the front of the indicator leaflet with the red test defilement remains visible in the open part of the retainer. Place or hang the retainer with the inserted Wash-Checks into the washer-disinfector. The opening showing the red test defilement must face towards the inside of the washer-disinfector.
Note: for daily routine examination always insert the Wash-Checks with the normal load. Pay attention to the correct loading of the washer-disinfector (refer to manufacturer's specifications).
Evaluating the Indicator
At the end of the cycle remove the Wash-Checks from the washer-disinfector. Remove the indicator leaflet from the retainer and examine the cleaning result.
| untreated or | insufficient | optimal |
| critical result | cleaning result | cleaning result |
Documentation
Register the result of the indicator leaflet into the washer-disinfector’s logbook (order Nr. 840001) and discard the indicator.