Wash-Checks
Chemical control system to monitor the cleaning achieved with washer-disinfectors during daily routine usage.

Intended purpose
Easy too use control system to monitor the cleaning efficiency of medical instruments in washer-disinfectors.

 

Product Order Nr. Package size Storage Durability
Wash-Checks retainer 830001 1 Piece dark and dryat room temperature 2 years startingfrom date of manufacture
Wash-Checks 830010 10 Pieces
indicator leaflets 830050 50 Pieces
  830100 100 Pieces

 

 

 

 

 

 

 

Application/ Process
Insert the indicator leaflet laterally into the retainer, making certain that the front of the indicator leaflet with the red test defilement remains visible in the open part of the retainer. Place or hang the retainer with the inserted Wash-Checks into the washer-disinfector. The opening showing the red test defilement must face towards the inside of the washer-disinfector.

Note: for daily routine examination always insert the Wash-Checks with the normal load. Pay attention to the correct loading of the washer-disinfector (refer to manufacturer's specifications).

Evaluating the Indicator
At the end of the cycle remove the Wash-Checks from the washer-disinfector. Remove the indicator leaflet from the retainer and examine the cleaning result.

untreated or insufficient optimal
critical result cleaning result cleaning result


Documentation
Register the result of the indicator leaflet into the washer-disinfector’s logbook (order Nr. 840001) and discard the indicator.